65. Impact of Segmentation Resolution on Numerically Computed Electric Potential Around Biomedical Electrode in an Anisotropic and Dispersive Tissue
Keywords:
anisotropic dispersive tissue, finite difference method, computational accuracy, electric potential, quasi-static frequencyAbstract
Numerical simulations are vastly employed for field distribution computations in modelled conductive and dispersive media with different excitation types and parameters. The reliability of the numerical simulation becomes a crucial part when applied to human tissue. This paper presents a simulation of current-injection technique applied on a modelled muscle tissue with anisotropic and dispersive properties. The source is modelled as a point-current source at a quasi-static frequency, represented with a finite radius allowing for exact error computations. The numerical values obtained by employing the frequency-domain 7-point stencil finite difference method (FMD7) are compared with the analytically obtained results, for various nodal distances. The computed errors are used to empirically find the maximal nodal distances, and the adequate nodal distance ratios. The analysis is expanded on the 19-point stencil FDM (FDM19). The results indicate that the impact of nodal distances is similar for both FDM7 and FDM19.Downloads
Published
2026-09-24
Issue
Section
Papers