15. Development and Validation of a Laboratory Procedure for Oscilloscope Calibration
Keywords:
oscilloscope, calibration, measurement uncertainty, EURAMET cg-7, bandwidthAbstract
This paper presents the development and validation of laboratory calibration procedures for digital storage oscilloscopes (DSO), covering four key metrological parameters: amplitude (vertical system), timebase (horizontal system), rise time, and bandwidth. For each parameter, a dedicated measurement model, GUM-compliant uncertainty budget, and representative numerical examples from real laboratory measurements on a GW Instek GDS1072B oscilloscope are provided. A comparative analysis of EURAMET cg-7 version 1.0 (2011) and version 2.0 (2025) highlights the critical methodological improvements for bandwidth calibration of DSOs, including the explicit treatment of the brick-wall frequency response, the numerical determination of the sensitivity coefficient Sr, and the decomposition of the bandwidth uncertainty budget into seven components. Prior to the main calibration, the Transmille SPC600 BW Mode output was verified using a Rohde & Schwarz RTO2034 (3 GHz) over 46 frequency points, confirming its suitability up to 600 MHz. All procedures were performed at the Electrical Measurements Laboratory (LEM) of FEIT, Ss. Cyril and Methodius University in Skopje.Downloads
Published
2026-09-24
Issue
Section
Papers